Login / Signup

A Proof-of-Concept of Label-Free Biosensing System for Food Allergy Diagnostics in Biophotonic Sensing Cells: Performance Comparison with ImmunoCAP.

Rocio L EspinosaMaría Fe Laguna HerasFátima FernándezBeatriz SantamariaFrancisco Javier SanzaMaria Victoria MaiglerJuan J Álvarez-MillánVíctor Canalejas-TejeroMiguel Holgado
Published in: Sensors (Basel, Switzerland) (2018)
Food allergy is a common disease worldwide with over 6% of the population (200⁻250 million people) suffering from any food allergy nowadays. The most dramatic increase seems to be happening in children and young people. Therefore, improvements in the diagnosis efficiency of these diseases are needed. Immunoglobulin type E (IgE) biomarker determination in human serum is a typical in vitro test for allergy identification. In this work, we used a novel biosensor based on label-free photonic transducers called BICELLs (Biophotonic Sensing Cells) for IgE detection. These BICELLs have a thin film of nitrocellulose over the sensing surface, they can be vertical optically interrogated, and are suitable for being integrated on a chip. The BICELLs sensing surface sizes used were 100 and 800 µm in diameter. We obtained calibration curves with IgE standards by immobilizating anti-IgE antibodies and identified with standard IgE calibrators in minute sample amounts (3 µL). The results, in similar assay format, were compared with commercially available ImmunoCAP®. The versatility of the interferometric nitrocellulose-based sensing surface was demonstrated since the limit of detections for BICELLs and ImmunoCAP® were 0.7 and 0.35 kU/L, respectively.
Keyphrases