Effect of non-rigid connector on the stress distribution of implant-tooth supported fixed prostheses using different implant length and diameter: a comparative 3D finite element study.
Ghada H NaguibAbou Bakr Hossam HashemAhmed AbougaziaHisham A MouslyOsama A QutubMohamed T HamedPublished in: Journal of prosthodontics : official journal of the American College of Prosthodontists (2023)
Implant- tooth FDPs with rigid connector design using the combination of implant diameter and length (5.7mm × 10mm) demonstrated a better design modality thus predicting a higher success rate and more longevity than using non rigid connection option. This article is protected by copyright. All rights reserved.