Long-term durability after surgical aortic valve replacement with the Trifecta and the Intuity valve-a comparative analysis.

Paul WernerIuliana CotiAlexandra KaiderJasmin GritschMarkus MachAlfred KocherGuenther LauferMartin Andreas
Published in: European journal of cardio-thoracic surgery : official journal of the European Association for Cardio-thoracic Surgery (2021)
SVD was significantly more frequent in patients receiving a Trifecta valve and its implantation was an independent risk factor for the occurrence of SVD and aortic valve re-interventions. This comparative analysis of 2 low-gradient bioprosthesis put the long-term durability of the Trifecta valve in question and need to be taken into consideration when performing bioprosthetic SAVR.
Keyphrases
  • aortic valve
  • aortic valve replacement
  • aortic stenosis
  • transcatheter aortic valve implantation
  • transcatheter aortic valve replacement
  • risk assessment
  • ejection fraction