A comparison of white-to-white measurements obtained by Anterion AS-OCT versus three optical devices in refractive surgery candidates.

Ramin SaloutiAli AziziHesam HashemianMasoud Khorrami-NejadQassim AlkhafajiKia SaloutiMaryam GhoreyshiMohammad Hossein Nowroozzadeh
Published in: European journal of ophthalmology (2024)
This study found weak agreement between Anterion and Pentacam AXL, IOLMaster 700, and Orbscan IIz devices regarding WTW distance measurements in refractive surgery candidates. Therefore, it is not recommended to use Anterion's WTW measurements interchangeably with the other three devices, particularly for phakic intraocular lens sizing.