Validation of non-invasive point of care blood content analysis using the TensorTip™ MTX device: a method comparison study.

Sjoerd ServaasSilke de VreedeRuben L SmeetsAn StroobantsLucas T van EijkIgnacio MalagonCornelis Slagt
Published in: Clinical chemistry and laboratory medicine (2023)
Non-invasive blood content analysis with the TensorTip MTX device is not equivalent to and did not correlate sufficiently with conventional laboratory analysis. None of the parameters measured showed results within the limits of allowable error. Therefore, the use of the TensorTip MTX is not recommended for perioperative care.