Login / Signup

Failure to post-dilate BeSmooth peripheral stents to adult vessel size diameters during benchside tests.

Raymond N HaddadDamien BonnetSophie-Guiti Malekzadeh-Milani
Published in: Cardiology in the young (2023)
In our benchmark tests, extreme shortening, severe balloon rupture, or unpredictable stent fracture patterns at small balloon diameters limits safe post-dilation of BeSmooth stents beyond 13 mm. BeSmooth stents are not ideal candidates for off-label stent interventions in smaller patients.
Keyphrases
  • end stage renal disease
  • chronic kidney disease
  • newly diagnosed
  • physical activity
  • prognostic factors
  • peritoneal dialysis
  • climate change
  • stress induced
  • patient reported outcomes
  • drug induced