Analytical performance evaluation of two automated urine chemistry analyzers using two levels of commercially available quality control material.

J A HokampK FreemanKendal E Harr
Published in: The Journal of small animal practice (2023)
VetScan SA had marked negative pH bias versus VetLab UA resulting in clinically significant, overly acidic results. Specific gravity, nitrite, and leukocyte test pads should not be used. Both instruments had excellent performance in normal quality control material. While blood, glucose, protein and bilirubin are correctly identified as present in abnormal quality control material, exact concentrations cannot be interpreted due to imprecision. Only semiquantitative results, not numerical values implying quantification, should be reported from urine test strips.