Investigation of the halo-artifact in 68Ga-PSMA-11-PET/MRI.
Thorsten HeußerPhilipp MannChristopher M RankMartin SchäferAntonia Dimitrakopoulou-StraussHeinz-Peter SchlemmerBoris A HadaschikKlaus KopkaPeter BachertMarc KachelrießMartin T FreitagPublished in: PloS one (2017)
Halo-artifacts are particularly observed for 68Ga-PSMA-11-PET/MRI due to 1) the biodistribution of the PSMA-11-tracer resulting in large OBRs for bladder and kidneys and 2) inaccurate scatter correction methods currently used in clinical routine, which tend to overestimate the scatter contribution. If not compensated for, 68Ga-PSMA-11 uptake pathologies may be masked by halo-artifacts leading to false-negative diagnoses. Reducing the maximum scatter fraction was found to efficiently suppress halo-artifacts.