Radiation exposure of a mobile 3D C-arm with large flat-panel detector for intraoperative imaging and navigation - an experimental study using an anthropomorphic Alderson phantom.

Yashar NaseriUlrich HubbeChristoph ScholzJohannes BrönnerMarie T KrügerJan-Helge Klingler
Published in: BMC medical imaging (2020)
The larger 3D C-arm offers the possibility to reduce radiation exposures for specific applications despite its larger flat-panel detector with a larger field of view. However, due to the considerably higher radiation exposure of the larger 3D C-arm during lumbar 3D scans, the smaller 3D C-arm is to be recommended for short-distance instrumentations (mono- and bilevel) from a radiation protection point of view. The larger 3D C-arm with its enlarged 3D image set might be used for long instrumentations of the lumbar spine. From a radiation protection perspective, the use of the respective 3D C-arm should be based on the presented data and the respective application.