Investigating the encrustation of reinforced ureteral stents by computational flow dynamic simulations.
Benoît VogtPublished in: World journal of urology (2023)
The analysis of fluid dynamics through the tested stents confirmed that encrustations are possible in regions of stagnant flow and showed that stent models with the smoothest possible surface are preferable. The discontinuous flow model provided results that are closer to the findings observed in the clinic and should be more often integrated into CFD simulations.