Accuracy of the Withings BPM Connect Device for Self-Blood Pressure Measurements in General Population - Validation According to the Association for the Advancement of Medical Instrumentation/European Society of Hypertension/International Organization for Standardization Universal Standard.

Jirar TopouchianParounak H ZelveianZoya HakobyanHeghine GharibyanRoland Asmar
Published in: Vascular health and risk management (2022)
The Withings BPM Connect oscillometric device for home BP measurement fulfilled all of the accuracy requirements of the AAMI/ESH/ISO Universal Standard (ISO 81060-2:2018) in general population and consequently can be recommended for home BP measurements.