A 1-m non-resonant inelastic x-ray scattering spectrometer at BL15U, Shanghai Synchrotron Radiation Facility.

Dong-Dong NiXu KangShuai YanXin-Chao HuangTao XiongDong-Xu LiangKe YangLin-Fan Zhu
Published in: The Review of scientific instruments (2018)
We report the design, construction, and commissioning of a spectrometer for non-resonant inelastic x-ray scattering study installed at BL15U, Shanghai Synchrotron Radiation Facility. It features a 1-m vertical scattering arm. An energy resolution of 1.3 eV is achieved based on the 1 m Rowland circle and the diced Si(555) crystal analyzer with a fixed Bragg angle of about 88.8°. The inelastic squared form factors of 21S + 21P of helium with respect to the momentum transfer were measured and compared with the accurate and reliable theoretical calculations in order to verify the spectrometer. Furthermore, the spectrometer is designed to work in the momentum transfer region of 0 Å-1 < q < 8.68 Å-1 and to initially focus on the non-resonant inelastic x-ray scattering studies on gaseous samples.