Biventricular versus Conduction System Pacing after Atrioventricular Node Ablation in Heart Failure Patients with Atrial Fibrillation.
Maja IvanovskiMiha MrakAnja Zupan MežnarDavid ŽižekPublished in: Journal of cardiovascular development and disease (2022)
Conduction system pacing (CSP) modalities, including His-bundle pacing (HBP) and left bundle branch pacing (LBBP), are increasingly used as alternatives to biventricular (BiV) pacing in heart failure (HF) patients scheduled for pace and ablate strategy. The aim of the study was to compare clinical outcomes of HF patients with refractory AF who received either BiV pacing or CSP in conjunction with atrio-ventricular node ablation (AVNA). Fifty consecutive patients (male 48%, age 70 years (IQR 9), left ventricular ejection fraction (LVEF) 39% (IQR 12)) were retrospectively analysed. Thirteen patients (26%) received BiV pacing, 27 patients (54%) HBP and 10 patients (20%) LBBP. All groups had similar baseline characteristics and acute success rate. While New York Heart. Association (NYHA) class improved in both HBP ( p < 0.001) and LBBP ( p = 0.008), it did not improve in BiV group ( p = 0.096). At follow-up, LVEF increased in HBP (form 39% (IQR 15) to 49% (IQR 16), p < 0.001) and LBBP (from 28% (IQR 13) to 40% (IQR 13), p = 0.041), but did not change in BiV group ( p = 0.916). Conduction system pacing modalities showed superior symptomatic and echocardiographic improvement compared to BiV pacing after AVNA. With more stable pacing parameters, LBBP could present a more feasible pacing option compared to HBP.