Impact of prescription isodose level and collimator selection on dose homogeneity and plan quality in robotic radiosurgery.

Alexandra HellerbachMarkus EichnerDaniel RueßKlaus LuykenMauritius HoevelsMichael JudgeChristian BauesMaximilian RugeMartin KocherHarald Treuer
Published in: Strahlentherapie und Onkologie : Organ der Deutschen Rontgengesellschaft ... [et al] (2021)
Inhomogeneous dose distributions with PIL ≤ 70% can be used to minimize dose to normal tissue. PIL ≥ 90% is associated with a marked and significant increase in off-target dose exposure. Careful selection of collimators during planning is even more important.
Keyphrases
  • small cell lung cancer
  • quality improvement
  • robot assisted