Login / Signup

Defining the Parameters for Endovenous Microwave Ablation to Achieve Equivalence With Endovenous Laser Ablation, Using the Porcine Liver Model.

Anjali BachettaSimon CheungEmma R MooreDanny NguyenMelissa J KielyMark Steven Whiteley
Published in: Vascular and endovascular surgery (2024)
We have identified parameters for EMWA that gave equivalent tissue ablation in the porcine liver model to commonly used parameters used for EVLA. These need to be validated clinically, but as the model used has already been validated against clinical outcome in endovenous thermal ablation, there is little reason to suspect that these results would not be valid. As the power during EMWA is higher than EVLA, EVMA LEEDs are approximately 4-6 times higher than EVLA LEEDs to achieve the same thermal effect on the tissues.
Keyphrases
  • radiofrequency ablation
  • catheter ablation
  • gene expression
  • high resolution