Quantitative and Multiplexed Chopper-Based Time-Gated Imaging for Bioanalysis on a Smartphone.

Sahil S KananiHsin-Yun TsaiW Russ Algar
Published in: Analytical chemistry (2023)
Smartphones are emerging platforms for point-of-care diagnostics (POCDs), where the on-board camera is, for example, used to image fluorescence. Many laboratory instruments are capable of time-gated (TG) photoluminescence (PL) measurements─an analytical method leveraged by multiple commercial assay kits. When paired with long-lived PL emitters such as luminescent lanthanide complexes (LLCs), time-gating eliminates background from sample autofluorescence and many other sources. This capability is amenable to minimally processed samples and would thus be useful for POCDs on a smartphone-based platform. Here, we report a double-chopper design for TG PL imaging using a portable, 3D-printed, smartphone-based device. The rotation speed, dimensions, and overlap of the chopper blades and gaps set the timing parameters, with delay times on the order of hundreds of microseconds to milliseconds. The device was capable of quantitative TG imaging of PL from terbium(III) and europium(III) LLCs, including rejection of short-lived PL background from serum and tissue phantoms, spectral and temporal multiplexing, a model time-gated Förster resonance energy transfer (TG-FRET) assay, and imaging of cells. As the first smartphone-based demonstrations of these important analytical capabilities, this work is an important foundation for developing POCD methods based on TG PL imaging.