Login / Signup

Quantitative effects of acquisition duration and temporal resolution on the measurement accuracy of prostate dynamic contrast-enhanced MRI data: a phantom study.

Silvin Paul KnightJacinta Elizabeth BrowneJames Frances Mary MeaneyAndrew John Fagan
Published in: Magma (New York, N.Y.) (2017)
Errors in measured wash-in, wash-out, K trans, and v e parameters were minimised using T res ≤ 8.1 s and AD ≥ 360 s, with large errors recorded outside of this range.
Keyphrases