Purpose. To evaluate the short-term effect of standard automated perimetry (SAP) testing on intraocular pressure (IOP) in patients with open-angle glaucoma (OAG). Methods. We tested 45 patients (71 eyes) with OAG that had stable IOP under medical treatment. IOP was measured four times using an iCare rebound tonometer (RBT) immediately before, immediately after, 10 minutes after, and 20 minutes after SAP testing. Logistic regression analyses were performed to determine the relationships among SAP test duration, mean deviation of the SAP result, type of glaucoma medications, patient age, and significant IOP change (exceeding 2 mmHg) from baseline IOP. Results. The mean baseline IOP was 13.29 ± 3.06 mmHg. Although IOP changes immediately and 20 minutes after SAP testing were not statistically significant, the IOP change 10 minutes after SAP testing (-0.57 ± 1.84 mmHg) was statistically significant. However, the changes were within the margin of error of the RBT. Test duration, mean deviation, patient age, and type of glaucoma medications did not have a significant influence on IOP change (all P > 0.14). Conclusions. IOP measured by RBT did not vary significantly after SAP testing in patients with OAG. It may be not necessary to reject IOP measured after SAP testing in patients with OAG.