Login / Signup

Mechanically Sheared Axially Swept Light-Sheet Microscopy.

Jinlong LinDushyant MehraZach MarinXiaoding WangHazel M BorgesQionghua ShenSeweryn GałeckiJohn HaugKevin M Dean
Published in: bioRxiv : the preprint server for biology (2024)
We present a mechanically sheared image acquisition format for upright and open-top light-sheet microscopes that automatically places data in its proper spatial context. This approach, which reduces computational post-processing and eliminates unnecessary interpolation or duplication of the data, is demonstrated on an upright variant of Axially Swept Light-Sheet Microscopy (ASLM) that achieves a field of view, measuring 774 x 435 microns, that is 3.2-fold larger than previous models and a raw and isotropic resolution of ∼420 nm. Combined, we demonstrate the power of this approach by imaging sub-diffraction beads, cleared biological tissues, and expanded specimens.
Keyphrases
  • optical coherence tomography
  • high resolution
  • single molecule
  • electronic health record
  • big data
  • high throughput
  • high speed
  • gene expression
  • minimally invasive
  • deep learning
  • machine learning