Experimental comparison of four nonlinear magnetic detection methods and considerations on clinical usability.

Melissa Mathilde Horstman-van de LoosdrechtL AbelmannB Ten Haken
Published in: Biomedical physics & engineering express (2020)
Superparamagnetic iron oxide nanoparticles (SPIONs) are promising for clinical applications, because they have a characteristic nonlinear magnetic response when an external magnetic field is applied. This nonlinearity enables the distinct detection of SPIONs and makes measurements less sensitive to the human body and surgical steel instruments. In clinical applications, only a limited field strength for the magnetic detection is allowed. The signal to noise ratios (SNRs) of four nonlinear magnetic detection methods are compared. These methods include differential magnetometry and three variations of magnetic particle spectroscopy: frequency mixing, second harmonic detection and third harmonic detection. All methods were implemented on the same hardware and experimentally compared for various field strengths. To make the comparison fair, the same power was supplied to the excitation coil each time. In general, the SNR increases with increasing field strength. The SNR per drive field of all methods stabilizes or even decreases for field strengths above 6 mT. The second harmonic detection has the best SNR and the most room for improvement.
Keyphrases
  • loop mediated isothermal amplification
  • real time pcr
  • label free
  • endothelial cells
  • iron oxide nanoparticles
  • healthcare
  • high resolution
  • social media
  • patient reported outcomes
  • energy transfer