Novel Methods to Understand the Temporal Nature and Accuracy of Delivery for Insulin Infusion Pumps.
S Siddharth KumarDeval KariaArjun GopkumarPavan G KotyManish AroraPublished in: Journal of diabetes science and technology (2022)
The methods presented in this article help understand the delivery dynamics of liquid drug delivery devices. Our results indicate that a significant fraction of insulin delivery happens after the planned bolus duration, which might be important consideration for design of AP systems.