Login / Signup

Quantitative analysis of treatments using real-time image gated spot-scanning with synchrotron-based proton beam therapy system log data.

Takaaki YoshimuraShinichi ShimizuTakayuki HashimotoKentaro NishiokaNorio KatohHiroshi TaguchiKoichi YasudaTaeko MatsuuraSeishin TakaoMasaya TamuraSodai TanakaYuichiro AbeYuto MatsuoHiroshi TamuraKenji HoritaKikuo UmegakiHiroki Shirato
Published in: Journal of applied clinical medical physics (2020)
A synchrotron-based real-time image gated spot-scanning proton beam therapy (RGPT) system with inserted fiducial markers can irradiate a moving tumor with high accuracy. As gated treatments increase the beam delivery time, this study aimed to investigate the frequency of intra-field adjustments corresponding to the baseline shift or drift and the beam delivery efficiency of a synchrotron-based RGPT system. Data from 118 patients corresponding to 127 treatment plans and 2810 sessions between October 2016 and March 2019 were collected. We quantitatively analyzed the proton beam delivery time, the difference between the ideal beam delivery time based on a simulated synchrotron magnetic excitation pattern and the actual treatment beam delivery time, frequency corresponding to the baseline shift or drift, and the gating efficiency of the synchrotron-based RGPT system according to the proton beam delivery machine log data. The mean actual beam delivery time was 7.1 min, and the simulated beam delivery time in an ideal environment with the same treatment plan was 2.9 min. The average difference between the actual and simulated beam delivery time per session was 4.3 min. The average frequency of intra-field adjustments corresponding to baseline shift or drift and beam delivery efficiency were 21.7% and 61.8%, respectively. Based on our clinical experience with a synchrotron-based RGPT system, we determined the frequency corresponding to baseline shift or drift and the beam delivery efficiency using the beam delivery machine log data. To maintain treatment accuracy within ± 2.0 mm, intra-field adjustments corresponding to baseline shift or drift were required in approximately 20% of cases. Further improvements in beam delivery efficiency may be realized by shortening the beam delivery time.
Keyphrases
  • electron microscopy
  • end stage renal disease
  • chronic kidney disease
  • newly diagnosed
  • big data
  • ejection fraction
  • machine learning
  • bone marrow
  • mesenchymal stem cells
  • data analysis
  • high intensity