Development of image quality related reference doses called acceptable quality doses (AQD) in paediatric CT exams in Qatar.

Mohammad Hassan KharitaHuda AlNaemiVishwanatha KiniShady AlkhazzamMadan M Rehani
Published in: European radiology (2020)
• The first study to integrate image quality assessment with analysis of patient dose indices shows feasibility for routine practice in other centres. • The values of acceptable quality dose (AQD) were provided for head, chest and abdomen CT of children divided into weight groups rather than age. They shall act as reference values for future studies. • Verification of our findings on proportional increase in exposure parameters (CTDIvol and DLP) with weight by other investigators shall be helpful.