In Vitro Quantification of Gutter Formation and Chimney Graft Compression in Chimney EVAR Stent-Graft Configurations Using Electrocardiography-Gated Computed Tomography.

Simon P OvereemEsmé J DonselaarJorrit T BoersenErik Groot JebbinkCornelis H SlumpJean-Paul P M de VriesMichel M P J Reijnen
Published in: Journal of endovascular therapy : an official journal of the International Society of Endovascular Specialists (2018)
Gutter volumes and stent compression are dynamic phenomena that reshape during the cardiac cycle. Compelling differences were observed during the cardiac cycle in all configurations, with the self-expanding (Endurant-Viabahn) chimney EVAR configurations having smaller gutters and less variation in gutter volume during the cardiac cycle yet more stent compression without affecting the chimney graft surface.