Improving the Bias of Comparator Methods in Analytical Performance Assessments Through Recalibration.
Stefan PleusManuel M EichenlaubThomas GerberElisabet Eriksson BoijaKonstantinos MakrisCornelia HaugGuido FreckmannPublished in: Journal of diabetes science and technology (2022)
Recalibration did lead to a decrease in bias and thus can reduce the impact of the choice of comparator method. The procedure should be verified in a prospectively designed setting.