NORMAL VALUES OF ESOPHAGEAL HIGH-RESOLUTION MANOMETRY: A BRAZILIAN MULTICENTER STUDY.

Gerson Ricardo DominguesNelson Henrique MichelsohnRicardo Guilherme ViebigDécio ChinzonAry NasiCarla Granja AndradeEponina Maria de Oliveira LemmeLuiz João Abrahão-JúniorMauricio Gustavo BravimMiguel Ângelo Nobre-E-SouzaNayara Salgado CarvalhoPaulo Jose Pereira Campos CarvalhoTomás Navarro RodriguesJoaquim Prado P de Moraes-Filho
Published in: Arquivos de gastroenterologia (2020)
This is the first report of normative data for the 24-channel water-perfused system in supine posture. It revealed higher integrated relaxation pressure and distal latency duration which suggest the need to change CC 3.0 cutoffs for this system. It is observed that there is a tendency that DCI >7000 mmHg.cm.s may represent the lower limit of hypercontractility, and when <700 mmHg.cm.s (<5% percentile) interpreted as ineffective esophageal motility or failcontraction. Also compared to Chicago 3.0, higher integrated relaxation pressure and duration of distal latency were found. We emphasize that these data must be confirmed by future studies.