Proposed Diagnostic Reference Levels in the Missouri/Southern Illinois Region Associated with Cone-beam Computed Tomography Use in Endodontics.

Arthur PopkowskiS Craig RhodesJohn F Hatton
Published in: Journal of endodontics (2024)
The DI1 and DAP values strongly correlated when 3 outlier CBCT machines (J Morita Veraview X800) using a novel 360° (full rotation) acquisition mode were excluded. The importance of selectable exposure parameters as directly related to ionizing radiation output is illustrated among the CBCT units surveyed. Although the actual FOV that is selected is ultimately dictated by the specific clinical requirements, a 4-cm × 4-cm FOV is recommended for specialist endodontics practice, whenever clinically practical, based on the decreased ionizing radiation output, as compared to that from a 5-cm × 5-cm FOV.