Login / Signup

Are serum cobalt and chromium levels predictors for patient-reported outcome measures in the ASR hip resurfacing arthroplasty?

Maarten C KoperReinier W A SpekMax ReijmanEline M van EsSara J BaartJan A N VerhaarP K Bos
Published in: The bone & joint journal (2023)
This study shows that increasing serum Co and Cr levels measured in patients with an ASR-HRA are predictive for deterioration in HHS and HOOS subscales in the following year. Increasing serum Co and Cr should forewarn both surgeon and patient that there is a heightened risk of failure. Continued and regular review of patients with an ASR-HRA implant by measurement of serum Co/Cr levels and PROMs remains essential.
Keyphrases
  • patient reported outcomes
  • gold nanoparticles
  • carbon nanotubes