Performance characteristics of the digital uMI550 PET/CT system according to the NEMA NU2-2018 standard.
Shuguang ChenPengcheng HuYushen GuHaojun YuHongcheng ShiPublished in: EJNMMI physics (2020)
NEMA NU-2 2018 testing was performed on the new SiPM-based uMI550 PET/CT system. The uMI550 shows a high-spatial resolution of less than 3 mm and a good timing resolution of 372 ps. It shows clinical significances on improving potentially diagnostic ability on small lesions.