Depth of interaction determination in monolithic scintillator with double side SiPM readout.

Matteo MorrocchiGiovanni AmbrosiMaria Giuseppina BisogniFilippo BosiMarco BorettoPiergiorgio CerelloMaria IonicaBen LiuFrancesco PennazioMaria Antonietta PilieroGiovanni PirroneVasile PostolacheRichard WheadonAlberto Del Guerra
Published in: EJNMMI physics (2017)
The two parameters allow to achieve a DOI resolution comparable with the state of the art, giving a continuous information about the three-dimensional interaction position of the scintillation. These results were obtained by using simple estimators and a detector scalable to a whole PET system. The DOI calibration obtained using lutetium natural radioactivity gives results comparable to the other standard method but appears more readily applicable to detectors mounted in a full PET scanner.