NiCoP/g-C 3 N 4 Nanocomposites-Based Electrochemical Immunosensor for Sensitive Detection of Procalcitonin.

Furong ChenLayue BaoYing ZhangRuili WangJinghai LiuWenfeng HaiYushuang Liu
Published in: Sensors (Basel, Switzerland) (2023)
Herein, an ultra-sensitive and facile electrochemical biosensor for procalcitonin (PCT) detection was developed based on NiCoP/g-C 3 N 4 nanocomposites. Firstly, NiCoP/g-C 3 N 4 nanocomposites were synthesized using hydrothermal methods and then functionalized on the electrode surface by π-π stacking. Afterward, the monoclonal antibody that can specifically capture the PCT was successfully linked onto the surface of the nanocomposites with a 1-(3-Dimethylaminopropyl)-3-ethylcarbodiimide hydrochloride (EDC) and N-Hydroxysuccinimide (NHS) condensation reaction. Finally, the modified sensor was employed for the electrochemical analysis of PCT using differential Pulse Voltammetry(DPV). Notably, the larger surface area of g-C 3 N 4 and the higher electron transfer capacity of NiCoP/g-C 3 N 4 endow this sensor with a wider detection range (1 ag/mL to 10 ng/mL) and an ultra-low limit of detection (0.6 ag/mL, S/N = 3). In addition, this strategy was also successfully applied to the detection of PCT in the diluted human serum sample, demonstrating that the developed immunosensors have the potential for application in clinical testing.
Keyphrases