Repeatability and inter-observer variation of choroidal thickness measurements using swept-source optical coherence tomography in myopic danish children aged 6-14 years.
Anders AsmussenBjarke S SmithFlemming MøllerTrine Møldrup JakobsenPublished in: Acta ophthalmologica (2021)
The 3DM was the most repeatable method in this paediatric population. It yielded a CR of 10.7 µm, confidence interval 2.4 µm, which makes the minimal detectable difference between two measurements 13.1 µm. Most inter-observer variation could be explained by the intramethod variation. Segmentation adjustment on 3D macula scans did not increase CR on a group level.