Development of a Matrix Sublimation Device with Controllable Crystallization Temperature for MALDI Mass Spectrometry Imaging.
Hanyi XieRi WuY L Winnie HungXiang-Feng ChenT-W Dominic ChanPublished in: Analytical chemistry (2021)
The size and distribution of matrix crystals deposited on the surface of a tissue section play a key role in the performance of MALDI mass spectrometry imaging (MALDI-MSI). In this study, uniform distribution and a restricted size of matrix crystals were achieved via a homemade matrix sublimation device with controllable crystallization temperature. The crystallization temperature was stably controlled at a subzero temperature, and homogeneous matrix crystals with diameters <0.2 μm were generated on the sample surface. Typical MALDI-MSI experiments of endogenous and exogenous components in the tissues of strawberries, kidneys, and mussels were conducted to examine the performance of the sublimator. Good reproducibility of MALDI-MSI was achieved, and the quality of ion images was significantly improved. These results demonstrate that the developed sublimator should have potential in matrix deposition for further high resolution MALDI-MSI application.