Login / Signup

Focused ultrasound heating in brain tissue/skull phantoms with 1 MHz single-element transducer.

Anastasia AntoniouNikolas EvripidouChristakis Damianou
Published in: Journal of ultrasound (2023)
The ABS skull demonstrated poorer performance in terms of tFUS compared to the Resin skull owing to its higher ultrasonic attenuation and porosity. The thin Resin phantom of 1 mm thickness provided an efficient acoustic window for delivering tFUS and heating up deep phantom areas. The results of such studies could be particularly useful for accelerating the establishment of a wider range of tFUS applications.
Keyphrases
  • image quality
  • resting state
  • optical coherence tomography
  • dual energy
  • functional connectivity
  • multiple sclerosis
  • magnetic resonance imaging
  • case control