Login / Signup

Accuracy and precision of ScvO2 measured with the CeVOX-device: A prospective study in patients with a wide variation of ScvO2-values.

Alexander HernerBernhard HallerUlrich MayrSebastian RaschLea OffmanRoland SchmidWolfgang Huber
Published in: PloS one (2018)
These findings suggest that the CeVOX device is capable to detect stability or instability of ScvO2_BGA. ScvO2_CeVOX accurately estimates ScvO2_BGA in case of stable values. However, intermittent measurement of ScvO2_BGA and re-calibration should be performed in case of substantial changes in ScvO2_CeVOX compared to baseline. Therefore, continuous measurement of ScvO2 with the CeVOX cannot replace ScvO2_BGA in instable patients. On the other hand, CeVOX might be useful for the monitoring of stable patients as a pre-test tool for more differentiated monitoring in case of changes in ScvO2_CeVOX.
Keyphrases
  • end stage renal disease
  • ejection fraction
  • newly diagnosed
  • chronic kidney disease
  • peritoneal dialysis
  • prognostic factors
  • patient reported outcomes
  • patient reported