3D fast low-angle shot (FLASH) technique for 3T contrast-enhanced brain MRI in the inpatient and emergency setting: comparison with 3D magnetization-prepared rapid gradient echo (MPRAGE) technique.
Sohil H PatelPrem P BatchalaKellan SchallertJames T PatrieSalma O AbbasDavid A OrnanSugoto MukherjeeThomas HuertaJohn P MuglerPublished in: Neuroradiology (2020)
On 3T brain MRI in the inpatient and emergency department setting, 1-min 3D-FLASH pulse sequence achieved comparable diagnostic performance to 4.5 min 3D-MPRAGE pulse sequence for detecting enhancing intracranial lesions, with reduced susceptibility and motion artifacts.