On the accuracy of bulk synthetic CT for MR-guided online adaptive radiotherapy.
Davide CusumanoLorenzo PlacidiStefania TeodoliLuca BoldriniFrancesca GrecoSilvia LongoFrancesco CelliniNicola DinapoliVincenzo ValentiniMarco De SpiritoLuigi AzarioPublished in: La Radiologia medica (2019)
Good agreement was found between dose calculated on sCTs and pCT (gamma passing rate 1%/1 mm equal to 91.2% ± 6.9% for sCTICRU and 93.7% ± 5.3% b or sCTtailor). The mean difference in estimating V95 (PTV) was equal to 0.2% using sCTtailor and 1.2% using sCTICRU, respect to pCT values CONCLUSIONS: The bulk sCT guarantees a high level of dose calculation accuracy also in presence of magnetic field, making this approach suitable to MRgRT. This accuracy can be improved by using patient-specific RED values.