Accuracy of flat panel detector CT with integrated navigational software with and without MR fusion for single-pass needle placement.

Marc C MabraySanjit DattaPrasheel V LillaneyTeri MooreSonja GehrischJason F TalbottMichael R LevittBasavaraj V GhodkePaul S LarsonDaniel L Cooke
Published in: Journal of neurointerventional surgery (2015)
Single pass needle placement in a DBS phantom using FDCT guidance is associated with a radial error of approximately 2.5-3.0 mm at a depth of approximately 80 mm. This system could accurately target sub-centimetre intracranial lesions defined on MR.