Impact of the blooming artefact on dental implant dimensions in 13 cone-beam computed tomography devices.
Victor Aquino WanderleyKarla de Faria VasconcelosAndre Ferreira LeiteRuben PauwelsSohaib ShujaatReinhilde JacobsMatheus Lima de OliveiraPublished in: International journal of implant dentistry (2021)
The visualization of the implant dimensional alteration differed between CBCT devices and scanning protocols with an increase in diameter ranging from 0.27 to 1.04 mm. For most CBCT devices, B&C adjustments allowed to reduce visualization of implant blooming.