Login / Signup

Hydrodynamic Resistance of Intracranial Flow-Diverter Stents: Measurement Description and Data Evaluation.

Benjamin CsippaDániel GyürkiGábor ZávodszkyIstván SzikoraGyörgy Paál
Published in: Cardiovascular engineering and technology (2019)
Based on our evaluation method a confidence band can be determined for a given deployment scenario. Collectively analysing the hydrodynamic resistance and the geometric indices, a deeper understanding of an implantation can be obtained. Our results suggest that to correctly interpret the hydrodynamic resistance of a scenario, the deployment length has to be considered. To demonstrate the applicability of the measurement, as a pilot study the results of four intracranial flow diverter stents of two types and sizes have been reported in this work. The results of these measurements even on this small sample size provide valuable information on differences between stent types and deployment scenarios.
Keyphrases
  • internal carotid artery
  • optic nerve
  • machine learning
  • clinical evaluation