An 8-channel Tx/Rx dipole array combined with 16 Rx loops for high-resolution functional cardiac imaging at 7 T.

Bart R SteensmaIngmar J VoogtTim LeinerPeter R LuijtenJesse HabetsDennis W J KlompCornelis A T van den BergAlexander J E Raaijmakers
Published in: Magma (New York, N.Y.) (2017)
This work demonstrates the performance of a combined loop/dipole array for cardiac imaging at 7 T. With this array, acceleration factors of 6 are possible without increasing the average g-factor in the heart beyond 1.27. Combining loops and dipoles in receive mode enhances the SNR compared to receiving with loops or dipoles only.