Changes in the Bronchial Cuff Pressure of Left-Sided Double-Lumen Endotracheal Tube by Lateral Positioning: A Prospective Observational Study.
Jong-Hae KimEugene KimIn-Young KimEun-Joo ChoiSung Hye ByunPublished in: Journal of clinical medicine (2021)
Proper bronchial cuff pressure (BCP) is important when using a double-lumen endotracheal tube (DLT), especially in thoracic surgery. As positional change during endotracheal tube placement could alter cuff pressure, we aim to evaluate the change in BCP of DLT from the supine to the lateral decubitus position during thoracic surgery. A total of 69 patients aged 18-70 years who underwent elective lung surgery were recruited. BCP was measured at a series of time points in the supine and lateral decubitus positions after confirming the DLT placement. The primary outcome was change in the initial established BCP (BCPi), which is the maximum pressure at which the BCP did not exceed 40 cmH2O without air leak in the supine position, after lateral decubitus positioning. As the primary outcome, the BCPi increased from 25.4 ± 9.0 cmH2O in the supine position to 29.1 ± 12.2 cmH2O in the lateral decubitus position (p < 0.001). Out of the 69 participants, 43 and 26 patients underwent surgery in the left-lateral decubitus position (LLD group) and the right-lateral decubitus position (RLD group) respectively. In the LLD group, the BCPi increased significantly (p < 0.001) after lateral positioning and the beginning of surgery and the difference value, ∆BCPi, from supine to lateral position was significantly higher in the LLD group than in the RLD group (p = 0.034). Positional change from supine to lateral decubitus could increase the BCPi of DLT and the increase was significantly greater in LLD that in RLD.