Evaluation of basic characteristics of 3-mm dose equivalent measuring instrument for evaluating lens exposure dose in radiotherapy.
Yutaro MoriTomonori IsobeHideyuki TakeiShohei MiyazakiSatoshi KamizawaTetsuya TomitaDaisuke KobayashiHideyuki SakuraiTakeji SakaePublished in: Journal of medical radiation sciences (2023)
We found that the characteristics of the 3-mm dose equivalent dosimeter in a high-energy radiation satisfy the IEC standards and have the same measurement accuracy as diagnostic areas such as Interventional Radiology.