Comparison of ultra-performance liquid chromatography and ARK immunoassay for therapeutic drug monitoring of voriconazole.

Diego Peña-LorenzoNoemi RebolloJosé Germán Sánchez-HernándezAranzazu Zarzuelo-Castañeda
Published in: Annals of clinical biochemistry (2021)
Although both techniques can be used for TDM of voriconazole, when a value close to the lower limit of the therapeutic range is determined by the ARKTM immunoassay, it would be better to verify the result by a non-automated technique to avoid possible underdosing.