An ingestible, battery-free, tissue-adhering robotic interface for non-invasive and chronic electrostimulation of the gut.
Kewang NanKiwan WongDengfeng LiBinbin YingJames C McRaeVivian R FeigShubing WangNingjie DuYuelong LiangQijiang MaoEnjie ZhouYonglin ChenLei SangKuanming YaoJingkun ZhouJian LiJoshua JenkinsKeiko IshidaJohannes KuosmanenWiam Abdalla Mohammed MadaniAlison HaywardKhalil B RamadiXinge YuGiovanni TraversoPublished in: Nature communications (2024)
Ingestible electronics have the capacity to transform our ability to effectively diagnose and potentially treat a broad set of conditions. Current applications could be significantly enhanced by addressing poor electrode-tissue contact, lack of navigation, short dwell time, and limited battery life. Here we report the development of an ingestible, battery-free, and tissue-adhering robotic interface (IngRI) for non-invasive and chronic electrostimulation of the gut, which addresses challenges associated with contact, navigation, retention, and powering (C-N-R-P) faced by existing ingestibles. We show that near-field inductive coupling operating near 13.56 MHz was sufficient to power and modulate the IngRI to deliver therapeutically relevant electrostimulation, which can be further enhanced by a bio-inspired, hydrogel-enabled adhesive interface. In swine models, we demonstrated the electrical interaction of IngRI with the gastric mucosa by recording conductive signaling from the subcutaneous space. We further observed changes in plasma ghrelin levels, the "hunger hormone," while IngRI was activated in vivo, demonstrating its clinical potential in regulating appetite and treating other endocrine conditions. The results of this study suggest that concepts inspired by soft and wireless skin-interfacing electronic devices can be applied to ingestible electronics with potential clinical applications for evaluating and treating gastrointestinal conditions.