Dynamic Modulation of Device-Arterial Coupling to Determine Cardiac Output and Vascular Resistance.
Steven P KellerBrian Y ChangQing TanZhengyang ZhangAhmad El KaterjiElazer R EdelmanPublished in: Annals of biomedical engineering (2020)
Clinical adoption of mechanical circulatory support for shock is rapidly expanding. Achieving optimal therapeutic benefit requires metrics of state to guide titration and weaning of support. Using the transvalvular positioning of a percutaneous ventricular assist device (pVAD), device:heart interactions are leveraged to determine cardiac output (CO) and systemic vascular resistance (SVR) near-continuously without disrupting therapeutic function. An automated algorithm rapidly alternates between device support levels to dynamically modulate physiological response. Employing a two-element lumped parameter model of the vasculature, SVR and CO are quantified directly from measurements obtained by the pVAD without external calibration or invasive catheters. The approach was validated in an acute porcine model across a range of cardiac (CO = 3-10.6 L/min) and vascular (SVR = 501-1897 dyn s/cm5) states. Cardiac output calculations closely correlated (r = 0.82) to measurements obtained by the pulmonary artery catheter-based thermodilution method with a mean bias of 0.109 L/min and limits of agreement from - 1.67 to 1.89 L/min. SVR was also closely correlated (r = 0.86) to traditional catheter-based measurements with a mean bias of 62.1 dyn s/cm5 and limits of agreement from - 260 to 384 dyn s/cm5. Use of diagnostics integrated into therapeutic device function enables the potential for optimizing support to improve outcomes for cardiogenic shock.
Keyphrases
- left ventricular
- pulmonary artery
- coronary artery
- heart failure
- pulmonary arterial hypertension
- ultrasound guided
- liver failure
- intensive care unit
- minimally invasive
- type diabetes
- mechanical ventilation
- drug induced
- electronic health record
- extracorporeal membrane oxygenation
- adipose tissue
- molecular dynamics simulations
- room temperature
- aortic dissection
- catheter ablation