Login / Signup

Multiple-Input Multiple-Output (MIMO) MRI: Combining Parallel Excitation and Parallel Reception for Enhanced Imaging.

Xianglun MaoNicole L VikeThomas M TalavageJoseph V RispoliDavid J Love
Published in: IEEE transactions on computational imaging (2019)
Magnetic resonance imaging (MRI) plays a critical role in visualizing the structure and functions of the human body. In order to accelerate imaging time and improve image quality, radio-frequency (RF) coil receive arrays are commonly employed to acquire the magnetic resonance (MR) signal. Similarly, multiple transmit coils have been shown to accelerate and refine RF excitation. In this work, we investigate the optimization of total imaging time and image accuracy when considering both the transmit and receive coil arrays; we term this strategy multiple-input multiple-output (MIMO) MRI. Our RF pulse design method is modeled by minimizing the excitation errors while simultaneously maximizing the signal-to-noise ratio (SNR) of the reconstructed MR image. It further allows a key tradeoff between the two optimizers. Additionally, multiple acceleration factors, varying numbers of receive coils used, maximum excitation error tolerance, and different excitation patterns are simulated and analyzed in this model. For a given excitation pattern, our method is shown to improve the SNR by 18-130% under certain acceleration schemes, as compared to conventional parallel transmission methods, while simultaneously controlling the excitation error within a desired scope (NRMSE ≤ 0.12).
Keyphrases