Integration of an MC-80 Digital Image Analyzer With an Automated BC-6800Plus Hematology Analyzer Enables Accurate Platelet Counting in Samples With EDTA-Induced Pseudothrombocytopenia.

Min-Kyung SoJungwon HuhSeunghwan KimSholhui Park
Published in: Annals of laboratory medicine (2024)
BC-6800Plus PLT-O measurement results are close to the actual values obtained by platelet deaggregation with PTCP samples. Integrating the BC-6800Plus with a digital imaging analyzer effectively improved the diagnosis of PTCP and reduced the requirement for additional laboratory procedures.