Analytical effects on clumped isotope thermometry: Comparison of a common sample set analyzed using multiple instruments, types of standards, and standardization windows.

William F DeflieseAradhna Tripati
Published in: Rapid communications in mass spectrometry : RCM (2020)
Our results show that background effects are the largest factor potentially affecting Δ47 results, and there may be an improvement in interlaboratory precision using carbonate standards. Critically, all techniques used for standardizing Δ47 results converge on a common slope as long as background effects are properly corrected. The use of carbonate standards is recommended as a component of standardization procedures.