Login / Signup

Verification Study of Residual Activity Measurements After Yttrium-90 Radioembolization with Glass Microspheres.

Sander C EbbersB KunnenB J van NieropJ L M BemelmansG C KrijgerM G E H LamA J A T Braat
Published in: Cardiovascular and interventional radiology (2020)
This study found a significant disagreement between the residual activities measured by the survey meter, compared to measurements by PET and dose calibrator. If relatively high amounts of residual activity are encountered using the exposure rate measurement with a survey meter, additional quantification should be considered using either PET/CT or a dose calibrator measurement.
Keyphrases
  • pet ct
  • positron emission tomography
  • liquid chromatography